home 

 

digiflume logo 

 

contacts 

 

download 

 

disclaimer 

 

 

 

 

 

 

Copyright © 2002 demondo
All rights reserved

 

digiflume logo

specification

Mechanical

Dimensions 110x200x365mm (wxdxh) excluding connectors, weight 3.1 kg.

Fluid connections. Input: 10-32 UNF female. Accepts Swagelok® SS-1F2-GC type or equivalent chromatography fitting. Output: 5/16"-24 UNF female. Accepts Swagelok® SS-1-100-OR (1/16") or SS-1-200-OR (1/8") tube connector.

Nominal barrel volume: 10, 20, 50 or 100 ml, maximum working pressure 700 bar (100 bar for 100 ml barrel).

Chemical compatibility: fluids compatible with Viton® elastomer, FEP, PEEK, PTFE polymers, AISI316 Stainless Steel.

Operating temperature +5 - +45 °C
Storage temperature 0 - +55 °C

Electrical

Power supply 8-12VDC @ 300 mA max (plug-top adapter), concentric connector, centre positive (2.5 mm diameter).

Data transmission: RS485 half duplex serial interface. Two connectors provided (one male and one female) to facilitate daisy-chaining of multiple devices.
Pin-out:

  1. Signal B, negative (-)
  2. Signal A, positive (+)
  3. 0V return
Accuracy, resolution and repeatability

Temperature: Resolution 0.1 °C, accuracy 0.5 °C

Dielectric constant: Resolution 0.1, accuracy ±2.5%of reading (Mode 1,3)(1)

Volume: Resolution: 1 µL, repeatability: ±0.05% FS (Mode 4)(3)
Accuracy:
Mode 1 ±2.5% of reading (1)
Mode 2 ±3% of reading (2)
Mode 3 ±2.5% of reading (1)
Mode 4 ±0.25% FS (3)

Figures are valid 15 minutes after power-up, and at a constant ambient temperature of 25 °C.


Notes:
1. Limited by mechanical tolerances of auxiliary cell (Ref. manual)
2. Dependent on rate-of-change of fluid interface position (Ref. manual)
3. For 0.1M KCl solution, within 30 days of calibration.
4. Specifications subject to change without notice.